Volumetric Karl Fischer Titrator
- Support KF Titration Mode and Titer Detection Mode.
- Features of auto-filling, auto-purging and auto-mixing of the reagents ensure safe handling of Karl Fischer chemicals.
- Automatic/manual background drift correction ensures accurate results.
- Selectable units including mg, mg/L, %, ppm, etc.
- Settable parameters, including measurement unit, polarization current,stirring rate, titration rate, stop volume, endpoint potential, stop criterionand etc.
- Store up to 200 titration data sets (GLP-compliant).
Coulometric Karl Fischer Titrator
- Generator cell with diaphragm.
- Automatic/manual background drift correction ensures accurate results.
- Selectable units including μg,mg,%,ppm,mg/L,μg/mL, etc.
- Settable parameters, including measurement unit, polarization current, stirring rate, generator current, endpoint potential, stop criterion and etc.
- Store up to 200 titration data sets (GLP-compliant).
GENERAL FEATURES
- LCD display screen.
- Titration methods, curves and results are displayed in detail.
- Volumetric KF Titration and Coulometric KF Titration are supported.
- Data analysis feature helps user review and compare results.
- Data can be easily transferred to printer via RS-232 communication interface.
- Reset feature automatically resumes all settings back to factory default options.
SPECIFICATIONS OF Karl Fischer Moisture Analyzer
Model |
WT-KF-40C |
|
Volumetric Titration |
Range |
0.1mg~250mg |
Resolution |
0.1mg |
|
Working Current |
1μA±0.2μA; 50μA±10μA |
|
Repeatability |
≤0.5% |
|
General |
Power Supply |
AC Adapter, 100-240 VAC input |
Dimensions |
341×400×400 mm |
|
|
Weight |
10kg |