Product description
1)oxidizing gas tester used to measure the oxidizing for a kind of gas or mixture gas, whether higher than the air oxidation, it can be used to test the toxicity and corrosive gases or mixtures of gases.
2) comply with ISO10156-2010 “gases and gas mixtures selection of cylinder valve exhaust port by measuring the fire potential and oxidizing ability” Part IV oxidation of gas and mixed gas, “chemical hazard classification test methods for gas and gas mixtures potential and oxidizing ability ” part 4 oxidative capacity of the combustion gases and gas mixtures.
Product Features
- Test part and control system independent design to ensure safe on instrument control unit and the laboratory personnel;
- Supporting with PC operating software;
- Support for remote ignition, ensure the safety of laboratory personnel;
- Equipped with high-precision pressure sensor that can accurately measure the explosion pressure, and control gas partial pressure ratio of the concentration;
- Equipped with a magnetic stirring device;
- Electronic micro-processor controlled ignition time;
- With auto cleaning function;
- With automatic pressure relief device, after the end of the trial, you can choose automatic pressure relief to release of the test tank pressure;
- With multiple explosion-proof function, to protect safety of personnel;
- With automatic fault detection and sound and light alarm function;
- Testing process fully automated, without human intervention, support parameter setting, data storage, viewing reports.
Product Specifications
Operating environment | 5 °C ~ 30 °C, <85% RH |
Test tank material | stainless steel 316L |
Test tank maximum pressure | ≥5000Kpa |
Test tank capacity | ≥0.005m³ |
Test tank accessories | magnetic stirrer, stirring speed adjustable |
Ignition system metal rod | diameter ≥3mm |
Nickel-chromium wire diameter | 0.05 ~ 0.2mm |
Ignition energy | 10 ~ 20J |
Explosion pressure sensor detection range | 0 ~ 4000kPa |
Explosion sensor pressure | ≥6000kPa |
Explosion sensor measurement accuracy | ± 0.1% FS |
Explosion sensor response time | ≤1ms |
Gas sensor detection range | 0 ~ 200kPa |
Gas sensor measurement accuracy | ± 0.1% FS |
Temperature range | 0 ~ 150 °C |
Temperature accuracy | ± 1 °C |
Host Size | 830mm × 650mm × 1205mm |
Electronic control box size | 320mm × 500mm × 505mm |
Weight | 280kg |
Power supply | AC220V ± 10%, 50Hz ± 3Hz |
Rated power | 400W |
Ignition power | 3000W |