Cold Trap is used to capture water vapor and harmful gases emission from vacuum drying oven and pressure reduced concentration device, improving efficiency of vacuum system, extending life-span of vacuum pump.

Advantages
- Can be used in drying system for capacitor, battery pole & battery cell.
- It also can be used as pre-freezing bath and low temperature bath.
- Digital display for better control of the bath temperature, which makes it easier to start the vacuum pump at right time.
- Stainless steel 304 liquid bath can be used to do water or ethanol cooling experiments. If equipped with glass condenser, it also can be used to deal with acid or organic solvents.
- Designed with drain valve for easy discharge of collected liquid.
Typical Application
Specifications
Model | UIC-2000H | UIC-2000L | UIC-5000H | UIC-5000L | |
Minimum temp. (°C) | -40 | -80 | -40 | -80 | |
Cooling type | Direct cooling or Indirect cooling | ||||
Ambient temp. (°C) | 5~25 | ||||
Relative humidity (%) | ≤70 | ||||
Power supply (V/ Hz) | 220-240V~,50Hz | ||||
Temperature display | Digital Display | ||||
Sensor index | Pt100 | ||||
Setting method | Arbitrary setting within the operating temperature range | ||||
Safety protection | Delay, overcurrent, overheat | ||||
Dehumidification capacity | Max.1kg (water system) | Max.1kg (water system) | Max.4kg (water system) | Max.4kg (water system) | |
Cooling capacity (W) | -10°C | 640 | – | 920 | – |
-20°C | 400 | – | 520 | – | |
-30°C | 200 | – | 280 | – | |
-40°C | 20 | 260 | 40 | 400 | |
-60°C | – | 110 | – | 200 | |
-80°C | – | 40 | – | 60 | |
Refrigerant | R404A | R404A, R23 | R404A | R404A, R23 | |
Machine power (W) | 819 | 1580 | 835 | 1602 | |
Machine current (A) | 3.7 | 7.2 | 3.8 | 7.3 | |
Cold trap cavity (mm) | Ф160×240 | Ф160×240 | Ф220×240 | Ф220×240 | |
Material of reservoir | 304 | 304 | 304 | 304 | |
* OD of inlet/outlet | 12mm(OD) | 12mm(OD) | 12mm(OD) | 12mm(OD) | |
Dimensions W × D × H (mm) | 365×590×891 | 439×528×1191 | 573×584×1052 | 505×730×1387 | |
Weight (kg) | 60 | 100 | 85 | 120 |
* Outlet diameter of inlet/outlet port can be customized.