Description
The world’s leading dual wavelength ultraviolet oxidation technology, high precision and high sensitivity. high-performance CPU, touch screen intelligent control, with offline analysis and on-line analysis of optional features, user-friendly design, to replace the UV lamp and tubing without opening machine, simple operation. Conforms to “Chinese Pharmacopoeia” 2010 version of Appendix VIII R WPU TOC determination to meet the pharmacopoeia requirements for instruments: 1. TOC = TC-TIC, 2. system suitability test, 3. detection sensitivity (equal to or less than 0.05mg / L)
Work principle
The total organic components of the sample are catalyzed into Carbon Dioxide under the use of UV and Titanium dioxide. Using conductivity-detecting method to detect the amount of Carbon Dioxide. The Total carbon is get under the use of oxidation reactor while detecting the sample in the positive electrode, and the Total inorganic Carbon is get without the use of oxidation reactor while detecting the sample in the negative electrode. And finally, the Total organic carbon equals the difference between the Total carbon and the Total inorganic carbon of the samples. : TOC=TC-TIC.
Features
- High accuracy and high sensitivity, easy to operate.
- Humane operating screen, owns one-key-run function, pipeline self-clean function.
- High-performance CPU, touch LCD screen design, ultra-clear 640*480 real color monitor.
- No need to disassemble the chassis when exchange the UN light and pump tube.
- Detecting up-limit can be set and self-alarm function start while surpass the limit.
- RS232 date interface, history date can be saved for 6 months.
- Integrative with offline detection and online detection.
- Attaching with a mini-printer interface (external printer)
Appliance
- Used for TOC detecting for purified water, injection water, deionized water in pharmaceutical industry and also the ultra-pure water in semiconductor industry.
- Used to validate the cleaning results in pharmaceutical and biochemical area’s validation processes.
- The online detecting function supplies the inspection of the pharmaceutical’s water system, semiconducting’s ultra-pure water system and chip technology , power plant’s deionized water .
Technical parameters
Model |
TOC-1000U |
TOC-1000UB |
Power supply |
AC220V±22V |
AC220V±22V |
Rate |
50HZ |
50HZ |
Power |
95W |
95W |
Dimension |
340*390*145mm |
400*260*295mm |
Weight |
15kg |
15kg |
TOC detecting scope |
0.001mg/L-1.00mg/L |
0.001mg/L-1.00mg/L |
Accuracy |
±5% |
±5% |
Detecting time |
≤4min/Time |
≤4min/Time |
Respond time |
≤15min |
≤15min |
Sample flow rate |
0.5ml/min |
0.5ml/min |
Temperature |
1 to 95°C |
1 to 95°C |
Humidity |
≤85%RH |
≤85%RH |
Repeated errors |
≤3% |
≤3% |
Zero point drifting |
±5% |
±5% |
Range drifting |
±5% |
±5% |
Linearity error |
±5% |
±5% |