Characteristics
- Tobe used in production of medicines, dye-stuffs, perfumes and other organic crystal
- UsePID & PWM to control the
- Automaticallyrecords the melting
- Calculateaverage value of initial and final melting points
- Equippedwith clock
- Canconnect to a computer for data
- Touch
Specifications
Parameters/Model | MP-2C | MP-2A | MP-1C | MP-1B | MP-2 |
Measuring range | RT ~ 360 ℃ | RT ~ 300 ℃ | RT ~ 400 ℃ | RT ~ 320 ℃ | RT ~ 300 ℃ |
Measuring mode | Auto | ||||
Measuring methods | 8 | ||||
Resolution | 0.1 ℃ | ||||
Linear heating-up rate | 0.1 ℃ /min ~20 ℃ /min | ||||
Accuracy | ≤200 ℃: ±0.4℃ ;>200 ℃: ±0.7℃ | ||||
Repeatability | 0.3 ℃(heating rate: 1.0℃/min) | ||||
Linear heating rate error | ±10% | ||||
Number of capillary | 3 | 1 | |||
Capillary dimensions | Outer diameter: φ1.4mm Inner diameter: φ1.0mm Length: 90mm | ||||
Height of sample | 3mm | ||||
Data storage | 1000 set | / | 1000 set | / | / |
Display | 7 inch touch LCD | ||||
Printer(Thermal) | Standard accessory | / | Optional accessory | / | / |
Interface | RS232,USB | ||||
Power supply | 100~240VAC,50/60Hz,100W | ||||
Package Dimension | 580mm×445mm×320mm | 500mm×445mm×330mm | |||