PT-D445-1005(1005A)
PT-D445-1005(1005A) Kinematical viscosity apparatus conforms the ASTM D445 Standard Test Method for Kinematic Viscosity of Transparent and Opaque Liquids (and Calculation of Dynamic Viscosity).
Availability date:
Overview
PT-D445-1005(1005A) Kinematical viscosity apparatus conforms the ASTM D445 Standard Test Method for Kinematic Viscosity of Transparent and Opaque Liquids (and Calculation of Dynamic Viscosity). It is used to test the of liquid petroleum products, both transparent and opaque, by measuring the time for a volume of liquid to flow under gravity through a calibrated glass capillary viscometer. The dynamic viscosity, η, can be obtained by multiplying the kinematic viscosity, ν, by the density, ρ, of the liquid. under certain constant temperature. The range of kinematic viscosities covered by this apparatus method is from 0.2 to 300 000 mm2/s at all temperatures
PT-D445-1005: temperature controlling precision,±0.1°C
PT-D445-1005A: temperature controlling precision,±0.01°C
Features
Technical parameters
Rated voltage | AC220V±10% 50Hz |
Temperature control range | 20~100°C |
Temperature control mode | Digital display temperature control |
Precision of temperature control | ±0.1°C(±0.01°C) |
Output power | 1800W |
Ambient requirement | Temperature: 20~30°C Humidity≤85% |
We can provide varies kinds of ASTM thermometers and capillary viscometers for this kinematic viscometer apparatus. See detailed information as below
Cannon-Fenske Opaque Viscometers
Size | Approx. | Kinematic Viscosity Range mm2/s(cSt) |
25* | 0.002 | 0.5 - 2 |
50 | 0.004 | 0.8 - 4 |
75 | 0.008 | 1.6-8 |
100 | 0.015 | 3-15 |
150 | 0.035 | 7-35 |
200 | 0.1 | 20-100 |
300 | 0.25 | 50-250 |
350 | 0.5 | 100-500 |
400 | 1.2 | 240-1200 |
450 | 2.5 | 500-2500 |
500 | 8 | 1600-8000 |
600 | 20 | 4000-20000 |
650 | 45 | 9000-45000 |
700 | 100 | 20000-100000 |
Cannon-Fenske Routine Viscometers
Size | Approx. | Kinematic Viscosity Range mm2/s(cSt) |
25* | 0.002 | 0.5 - 2 |
50 | 0.004 | 0.8 - 4 |
75 | 0.008 | 1.6-8 |
100 | 0.015 | 3-15 |
150 | 0.035 | 7-35 |
200 | 0.1 | 20-100 |
300 | 0.25 | 50-250 |
350 | 0.5 | 100-500 |
400 | 1.2 | 240-1200 |
450 | 2.5 | 500-2500 |
500 | 8 | 1600-8000 |
600 | 20 | 4000-20000 |
650 | 45 | 9000-45000 |
700 | 100 | 20000-100000 |
Ubbelohde Viscometers
Size | Approx. | Kinematic Viscosity Range mm2/s(cSt) |
0* | 0.001 | 0.3 -1 |
0C | 0.003 | 0.6 - 3 |
0B | 0.005 | 1-5 |
1 | 0.01 | 2-10 |
1C | 0.03 | 6-30 |
1B | 0.05 | 10-50 |
2 | 0.1 | 20-100 |
2C | 0.3 | 60-300 |
2B | 0.5 | 100-500 |
3 | 1.0 | 200-1000 |
3C | 3.0 | 600-3000 |
3B | 5.0 | 1000-5000 |
4 | 10 | 2000-10000 |
4C | 30 | 6000-30000 |
4B | 50 | 10000-50000 |
5 | 100 | 20000-100000 |
Thermometer
Item | Division Value(°C) | Thermometer Range(°C) | Size | Length(mm) | OD(mm) | Immersion (mm) |
28C* | 0.05 | 36.4-39.4 | Stick | 305±5 | 6.0-8.0 | Full |
29C | 0.05 | 52.6-55.4 | Stick | 305±5 | 6.0-8.0 | Full |
44C | 0.05 | 18.6-21.4 | Stick | 305±5 | 6.0-8.0 | Full |
45C | 0.05 | 23.6-26.4 | Stick | 305±5 | 6.0-8.0 | Full |
46C | 0.05 | 48.6-51.4 | Stick | 305±5 | 6.0-8.0 | Full |
47C | 0.05 | 58.6-61.4 | Stick | 305±5 | 6.0-8.0 | Full |
48C | 0.05 | 80.6-83.4 | Stick | 305±5 | 6.0-8.0 | Full |
110C | 0.05 | 133.6-163.4 | Stick | 305±5 | 6.0-8.0 | Full |
118C | 0.05 | 100-500 | Stick | 305±5 | 6.0-8.0 | Full |
120C | 0.05 | 38.6-41.4 | Stick | 305±5 | 6.0-8.0 | Full |
121C | 0.05 | 98.6-101.4 | Stick | 305±5 | 6.0-8.0 | Full |
129C | 0.05 | 91.6-94.4 | Stick | 305±5 | 6.0-8.0 | Full |
132C | 0.05 | 148.6-151.4 | Stick | 305±5 | 6.0-8.0 | Full |