ICP-AES650
ICP-AES650 Inductively Coupled Plasma Emission Spectrometer, also known as ICP Spectrometer ( Inductive Coupled Plasma )
Availability date:
Working principle of ICP-AES650 spectrometer
ICP-AES650 Inductively Coupled Plasma Emission Spectrometer, also known as ICP Spectrometer ( Inductive Coupled Plasma ), uses inductively coupled high-frequency plasma as the excitation light source, and uses the atomic or ion emission characteristic spectrum of each element to determine the composition of the substance. Qualitative and quantitative analysis of elements. Also known as ICP atomic emission spectrometer. The atomic emission spectrometry analysis process is mainly divided into three steps, namely excitation, spectroscopy and detection.
Description of ICP-AES650 spectrometer technical performance indicators
1. Radio frequency generator ( RF ):
circuit type | Inductive feedback self-excited oscillation circuit, fully automatic ignition, coaxial cable output matching and tuning, power feedback, and closed-loop automatic control. | working frequency | 40.68MHZ ± 0.05% |
frequency stability | <0.1% | ||
Output Power | 800-1500W | ||
output power stability | ≤ 0.3% | ||
Electromagnetic Field Leakage Intensity | from the fuselage30cm | ||
Electric field strength E | < 1 V/m | ||
Magnetic field strength H | <0.2A |
2. Sample injection system :
Output working coil inner diameter | 25mm, 3 turns | torch _ | Triple Concentric Quartz Torch |
Coaxial sprayer outer diameter | 6mm | Double cylinder mist outer diameter | 35mm |
Argon Flow Meter Specifications and Carrier Gas Pressure Gauge Specifications | Plasma gas flow meter ( 100-1000 ) L/h ; (1.6 -16L/min) | ||
Auxiliary air flow meter ( 10-100 ) L/h ; (0.16 -1.66L/min) | |||
Carrier gas flow meter ( 10-100 ) L/h ; (0.16 -1.66L/min) | |||
Carrier gas regulator valve ( 0 -0.6MPa ) |
3. Photomultiplier Tube Specifications:
Photomultiplier tube negative high voltage | 200-1000V Stability <0.05% |
Photomultiplier tube current measurement range | 10-12 -10-4A |
Signal acquisition is V/F exchange | 1mV corresponds to 100Hz |
4. Spectroscopic system :
light path | Czerny-Turner | focal length | 1000mm |
Grating Specifications | Ion etched holographic grating | Groove Density | 3600 lines /mm |
resolution | ≤ 0.008nm | Line area | 80 ×110mm |
Scan wavelength range | 3600 lines /mm | wavelength range | 195-800nm |
lens | ф 30 1 : 1 imaging | entrance slit | 20µm |
Transmitter Specifications | 78 * 105* 16 mm | exit slit | 20 µm |
Stepping motor drive minimum step | 0.0005nm |
ICP-AES650 Spectrometer application
1. Classification of detectable objects:
Including: metals ( steel , non-ferrous metals )
Chemicals, pharmaceuticals, petroleum, resins, ceramics, etc.
Biology, medicine, food, etc.
Environment ( tap water , environmental water , soil , atmospheric dust ) , etc.
2. Detection range
ICP atomic emission spectrometer is a large-scale analytical instrument that can determine the metal elements and some non-metal elements in the sample, and can perform qualitative, semi-quantitative and quantitative analysis on the sample. It can measure from micro to constant
3. ICP-AES650 spectrometer can detect element detection limit
4. The cost of using ICP spectrometer
The ICP atomic emission spectrometer needs to consume argon gas during use. The purity of argon gas that meets the requirements is 99.99%, and the consumption standard is about 12L/min.
ICP-AES650 spectrometer features
1. Fast analysis speed: Fastest analysis of more than 20 elements in one minute
2. High precision: Relative standard deviation RSD ≤ 1.5%
3. Good stability: Relative standard deviation RSD ≤ 2.0%
4. Low detection limit: μg/L
5. Many analysis elements: Quantitative analysis of more than 70 elements
6. Easy to operate: The new Windows operating environment is fully functional
7. Fully automatic ignition: Reduce the difficulty of use, convenient and quick
8. Multiple security protections: Low cooling water reminder protection, argon low value protection, arc extinguishing protection, triple protection is safer to use