Features
- Heating Circulator HRC series with function from high temperature to cooling process, from 300°C~50°C.
- Equipped with heating cooling container, with large heat transfer area, fast heat-up and cool-down, small heat transfer oil demand.
- Can heat-up and cool-down continuously.
- The whole cycle is full closed, there is no oil mist at high temperature, heat transfer oil cannot be oxidized and browning.
- With function of internal PT100 for correcting internal circulation.
- The whole system is closed with expansion tank, the expansion tank and liquid circulation are thermal insulation, do not participate in liquid circulation, regardless of high or low temperature, the medium in expansion tank is limited to 60°C.
Advantage
- Small space required
- Low filling Volume
- High Pump Capacity
- Rapid & efficient filling of the complete
- application with venting
- +60 °C max. expansion tank temperature
- Simple operation
- High level of safety through constant
- monitoring
Technical Parameters:
Model |
HRC-1820 |
HRC-5020 |
HRC-A020 |
HRC-3030 |
HRC-5030 |
HRC-A030 |
|
Temperature range °C |
50°C~170°C |
50°C~200°C |
50°C~200°C |
50°C~300°C |
50°C~300°C |
50°C~300°C |
|
Controller |
PID adaptive controller |
||||||
Temp. Control optional |
Heat-conducting medium outlet temp. control |
||||||
Communication protocol |
MODBUS RTU Protocol, RS 485 Interface |
||||||
Temperature feedback |
Heat-conducting medium temperature feedback PT100 |
||||||
Temp. accuracy |
±0.5°C |
±0.5°C |
±0.5°C |
±0.5°C |
±0.5°C |
±0.5°C |
|
Heating power |
2kW |
5.5kW |
10kW |
3kW |
5.5kW |
10kW |
|
Cooling |
300°C |
|
|
|
3kW |
5.5kW |
10kW |
200°C |
2kW at 170°C |
5.5kW |
10kW |
2.5kW |
5kW |
9kW |
|
100°C |
1.3kW |
4kW |
8kW |
1.3kW |
4kW |
7kW |
|
65°C |
0.6kW |
1.8kW |
3.4kW |
0.6kW |
1.8kW |
3kW |
|
|
There is a heat exchanger in the system, it needs to cool by filling water, open the water valve automatically as request. |
||||||
Circulation pump |
Max10L/min |
Max35L/min |
Max50L/min |
Max15L/min |
Max35L/min |
Max50L/min |
|
Heater |
Orient electric heater |
||||||
Independent Temperature limiter |
LABFREEZ |
||||||
Cooler |
KAORI Plate heat exchanger |
||||||
Operation panel |
Display setting temperature and testing temperature; Touch screen display |
||||||
Safety protection |
Self-diagnosis function; freezer overload protection; high pressure switch, overload relay, thermal protection device, |
||||||
Closed |
The whole system is full closed circulation, there is no oil mist at high temperature and no water vapor at low temperature,pressure do not rise up when system is running. The system will supplement oil automatically at low temperature. |
||||||
Connection size |
DN15 |
DN20 |
DN-20 |
DN-20 PN-10 Connection flange |
|||
Dimension |
320*420*500 |
500*550*800 |
500*550*800 |
500*550*800 |
600*600*1250 |
600*600*1250 |
|
Weight |
35kg |
78kg |
88kg |
85kg |
95kg |
105kg |
|
Power |
AC 220V 50HZ |
AC 380V 50HZ |
AC 380V 50HZ |
AC 380V 50HZ |
AC 380V 50HZ |
AC 380V 50HZ |
|
Case material |
Cold rolled steel |
||||||
Optional |
Optional 7-inch color touch screen controller, temperature curve record, data export to excel format. |
||||||
Optional |
SUS304 case material |
||||||
Optional power |
220V 60HZ three-phase, 440V~480V 60HZ three-phase |
Model “LY” means no water -cooling
Model |
HRC-A1520 |
HRC-A2520 |
HRC-A3820 |
HRC-A6020 |
HRC-A9520 |
HRC-A13020 |
HRC-A20020 |
|
Temperature range °C |
50°C~200°C |
|||||||
Controller |
PID adaptive controller |
|||||||
Temp. Control optional |
Heat-conducting medium outlet temp. control |
|||||||
Communication protocol |
MODBUS RTU Protocol, RS 485 Interface |
|||||||
Temperature feedback |
Heat-conducting medium temperature feedback PT100 |
|||||||
Temp. accuracy |
±0.5°C |
±0.5°C |
±1°C |
±1°C |
±1°C |
±1°C |
±2°C |
|
Heating power |
15kW |
25kW |
38kW |
60kW |
95kW |
130kW |
200kW |
|
Heater can be divided into groups. When the testing temperature reached to (setting temperature -5°C), |
||||||||
Cooling |
300°C |
|
|
|
|
|
|
|
200°C |
15kW |
25kW |
38kW |
60kW |
95kW |
130kW |
200kW |
|
100°C |
12kW |
20kW |
30kW |
48kW |
76kW |
100kW |
160kW |
|
65°C |
5kW |
8.5kW |
12kW |
19kW |
30kW |
40kW |
64kW |
|
|
There is a heat exchanger in the system and it needs water to cool. |
|||||||
Circulation pump |
Max50L/min |
Max110L/min |
Max110L/min |
Max250L/min |
Max250L/min |
Max400L/min |
Max400L/min |
|
Expansion tank volume |
35L |
65L |
90L |
160L |
240L |
300L |
500L |
|
Heater |
Orient electric heater |
|||||||
Independent temperature limiter |
LABFREEZ |
|||||||
Cooler |
Plate heat exchanger |
|||||||
Operation panel |
Display setting temperature and testing temperature; Touch key display |
|||||||
Safety protection |
Self-diagnosis function; freezer overload protection; high pressure switch, overload |
|||||||
Closed circulation system |
The whole system is full closed circulation, there is no oil mist at high temperature and no water vapor at low temperature, |
|||||||
Connection size |
DN-25 |
DN25 |
DN25 |
DN40 |
DN40 |
DN-50 |
DN-50 |
|
Dimension |
600*600*1250 |
600*600*1250 |
1200*750*1100 |
1200*750*1100 |
1400*750*1300 |
1400*750*1300 |
1500*850*1450 |
|
Weight |
100kg |
130kg |
170kg |
285kg |
318kg |
450kg |
685kg |
|
Power |
AC380V 50HZ |
AC380V 50HZ |
AC380V 50HZ |
AC380V 50HZ |
AC380V 50HZ |
AC380V 50HZ |
AC380V 50HZ |
|
Case material |
Cold rolled steel painting |
|||||||
Optional |
Optional 7-inch color touch screen controller, temperature curve record, data export to excel format. |
|||||||
Optional |
SUS 304 SUS 304 case material |
|||||||
Optional power |
220V 60HZ three-phase, 440V ~ 480V 60HZ three-phase |
Model |
HRC-A1530 |
HRC-A2530 |
HRC-A3830 |
HRC-A6030 |
|
Temperature range °C |
50°C ~ 300°C |
||||
Controller |
PID adaptive controller |
||||
Temp. Control optional |
Heat-conducting medium outlet temp. control |
||||
Communication protocol |
MODBUS RTU Protocol, RS 485 Interface |
||||
Temperature feedback |
Heat-conducting medium temperature feedback PT100 |
||||
Temp. accuracy |
±0.5°C |
±1°C |
±1°C |
±1°C |
|
Heating power |
15kW |
25kW |
38kW |
60kW |
|
Heater can be divided into groups. When the testing temperature reached to (setting temperature -5°C), |
|||||
Cooling |
300°C |
15kW |
25kW |
38kW |
60kW |
200°C |
13kW |
22kW |
32kW |
50kW |
|
100°C |
10kW |
17kW |
27kW |
40kW |
|
65°C |
4kW |
6kW |
11kW |
16kW |
|
|
There is a heat exchanger in the system and it needs water to cool. |
||||
Circulation pump |
Max50L/min |
Max110L/min |
Max110L/min |
Max250L/min |
|
Expansion tank volume |
50L |
80L |
120L |
220L |
|
Heater |
Orient electric heater |
||||
Independent temperature limiter |
LABFREEZ |
||||
Cooler |
Plate heat exchanger |
||||
Operation panel |
Display setting temperature and testing temperature; Touch screen display |
||||
Safety protection |
Self-diagnosis function; freezer overload protection; high pressure switch, overload |
||||
Closed |
The whole system is full closed circulation, there is no oil mist at high temperature and no water vapor at low temperature, |
||||
Connection size |
DN-25 |
DN-25 |
DN-25 |
DN-40 |
|
Dimension |
600*600*1250 |
600*600*1250 |
1200*750*1100 |
1200*750*1100 |
|
Weight |
115kg |
155kg |
195kg |
320kg |
|
Power |
AC380V 50HZ |
AC380V 50HZ |
AC380V 50HZ |
AC380V 50HZ |
|
Case material |
Cold rolled steel painting |
||||
Optional |
Optional 7-inch color touch screen controller, temperature curve record, data export to excel format. |
||||
Optional |
SUS 304 SUS 304 case material |
||||
Optional power |
220V 60HZ three-phase, 440V ~ 480V 60HZ three-phase |