FIT-1.2 Filter Integrity Tester

It is microcomputer control of automatic detection, with many kinds of test functions such as bubble point scanning, fast bubble point test and pressure holding method;

SKU: FIT-1.2 Category:

Main Characteristic
Online testing, does not interfere the downstream aseptic conditions
It has system self-check function, with an error report function for the system tightness and wrong operation;
It is microcomputer control of automatic detection, with many kinds of test functions such as bubble point scanning, fast bubble point test and pressure holding method;
Real-time print of test results;
It can be used to test the single core filter and multi-core filter.

Performance Parameter 

Model FIT-1.1 FIT-1.2
Power supply 170-240V AC, 50/60Hz;100W 170-240V AC, 50/60Hz;85w
Maximum operating pressure 8000mbar 8000 mbar
Minimum inlet pressure  100mbar   100 mbar
Dimension 400(W) x 380(D) x 100(H1) x 235(H2) 240(W) x 380(D1) x 280(D2) x 220(H1)
Measuring Range 500-6900mbar 500-6900mbar
Measuring Accuracy Net Volume Test:± 4%;Bubble Point:± 50mbar ;Diffusion Current:± 4% ; Sensitivity:± 1.0 mbar ;   Bubble Point:± 75 mbar
Operating condition Ambient Temperature:+5℃ ~ +35℃;Relative Humidity:10-80% Ambient Temperature:+5℃ ~ +35℃;Relative Humidity:10-80%
Test time consuming Net Volume Test:5 min±2min;    Diffusion Current Test:10 min±2min;
Simple bubble point test :15 min±2min;Enhanced Bubble Point Test:20 min±2min ;
Bubble point test :15 min±2min;   bubble point scanning : 20min±2min;Diffusion Current:10min±2min
Printing function printing, Output test parameters and test results printing, Output test parameters and test results
History recording function 500 group Test result( Including test curves ) Store 20 groups of test results
Display size:5.7″ TFT; homochromy size:73*39mm; homochromy
serial connection serial port :RS232
Language English English