CIT10-10S Aerosol spray enclosed space tester

CIT10-10S Aerosol spray enclosed space tester is used to assess flammability of the aerosol in a closed or restricted space. It is in line with the United Nations “on the Recommendations on the Transport of Dangerous Goods • Manual of Tests and Criteria” 31.5 enclosed space ignition test and “dangerous sprays enclosed space ignition test methods.”

SKU: CIT10-10S Category:

CIT10-10S aerosol spray enclosed space tester is used to assess flammability of the aerosol in a closed or restricted space. It is in line with the United Nations “on the Recommendations on the Transport of Dangerous Goods • Manual of Tests and Criteria” 31.5 enclosed space ignition test and “dangerous sprays enclosed space ignition test methods.”

Aerosol Spray Enclosed Space Tester consists of two parts, the display control section, the detection system composed of a cylindrical container body detection system, built-in light sensors and high-precision thermocouple sensor automatically determine whether knocking occurred. The operator can monitor the entire testing process through the display control section LCD screen.

 

Aerosol Spray Enclosed Space Tester

 

Product Features of Aerosol Spray Enclosed Space Tester

  • embedded processor, Windows CE operating system;
  • 8-inch LCD touch screen, display the test status in real time;
  • integration of the “standard model” and “Custom Mode” in two modes;
  • Spray automatic sprays and spray cans clamping device can automatically make the tank vertical and inverted;
  • pressure sensors, nanosecond light sensors double detection module, automatically detects whether a flame is burning or explosion in a closed space;
  • added to the ventilation system, the smoke exhaust from the barrel, to ensure test accuracy and safety of laboratory personnel;
  • testing process fully automated, without human intervention, support parameter setting, data storage, viewing reports, etc;
  • equipped with temperature and humidity module, real-time monitoring the temperature, humidity of test environment .

Aerosol Spray Enclosed Space Tester

Product parameters

MODEL CIT10-10S Aerosol spray enclosed space tester
Working environment  5 °C ~ 30 °C, <85% RH
Ambient temperature measuring range  -10 °C ~ + 60 °C
Ambient temp. measurement accuracy  ± 0.5 °C
Humidity measuring range  5 ~ 95% RH
Humidity measurement accuracy  ± 3.0% RH
combustion detection response time  ≤6ms
Differential pressure measurement range  -5 ~ 5kPa
Pressure measurement accuracy  ± 0.1% FS
Timing accuracy  ± 0.1s
ignition flame standard candle flame

adjust the pedestal height 20mm / a;

cylindrical vessel size  Ф600mm × 720mm
aerosol cans Location  aerosol tank automatically flips
compatible spray cans Dimensions  H 120 ~ 400mm, D 30 ~ 70mm
Distance from vessel to spray nozzle  30-35mm adjustable
effusion plate spray can accumulate discharged residual liquid

It can be withdrawn and processed at any time .