Total organic carbon tester

TOC-200

TOC (Total Organic Carbon), which directly shows the total organic content,is regarded as a key factorto evaluate pollution potentials of organic compound in water.patent laws, our TOC can meet requirements for application in water analysis, environment monitoring,pharmaceutical production, quality control, processing control etc.

More details

Product info

TOC (Total Organic Carbon), which directly shows the total organic content,is regarded as a key factorto evaluate pollution potentials of organic compound in water.patent laws, our TOC can meet requirements for application in water analysis, environment monitoring,pharmaceutical production, quality control, processing control etc.

So far, We launch 2 models TOC: TOC-200 (Catalytic combustion oxidation) and TOC-100 (Wetchemical oxidation by UV) which are widely used for quality control of drinking water, industrial water, sewage and waste water. Meanwhile, it has wide application in monitoring of river, lake, sea and surfacewater.

Application
Widely used in drinking water, industrial water, sewage, wastewater, and monitoring of rivers, lakes and oceans, surface water etc.

Main features:

  1. Signal management system with our own patent law which has great advantages of online setting, real-time monitoring, self testing and flow speed control to ensure high performance and safe operation.
  2. Low current system design also ensures operation safety.
  3. Different temperature setting for various samples ensures complete sample digestion so as to get more accurate measuring data.
  4. Adjust cooling module power according to sampling volume which improves drying performance to ensure dry gas into the detector.
  5. Automatic leak check system to avoid operation mistakes and improve instrument performance, so as to ensure operation safety and instrument’s safety
  6. TOC detector with 24 bits data solution extends monitoring range. Controlling system with 32bin processing technology greatly improves performance.
  7. TIC reaction pool heating, refrigeration, dehydration, liquid level monitoring of 4 in one, so that the air flow, liquid flow path is simple, save a lot of analysis time, while the detection limit to broaden;

Specifications: 


Model

TOC-200

TOC-100

Method

Catalytic combustion oxidation

Wet chemical oxidation by UV

Detector

NDIR

NDIR

Parameters

TC, TIC, TOC, DOC, POC

TC, TIC, TOC, DOC, NPOC

Digestion Mode

High Temperature Combustion

UV / persulfate oxidation

Operation Mode

PC Control

PC Control

Application

Liquid Sample, Gas Sample, Solid Sample

Liquid Sample, Gas Sample(Optional)

Gas Requirement

Oxygen, Nitrogen≥99.995%

Oxygen, Nitrogen≥99.995%

Measurement Range

0-30000mg/l

0-10000mg/l

Detection Limit

50μg/l

50μg/l

Repeatability

3%

3%

Maximum Salinity

85g/l

85g/l

Power

220±10V,50/60HZ, 1KW

220±10V,50/60HZ, 1KW

Size

430*455*440mm

430*455*440mm

 

10 other products in the same category: