Total Organic Carbon Analyzer (offline), With software and computer

TOC-1000S

TOC-1000S is a high-precision offline total organic carbon analysis instrument independently developed by our company.

More details

Product info

TOC-1000S is a high-precision offline total organic carbon analysis instrument independently developed by our company. The product uses ultraviolet oxidation and direct conductivity difference detection technology, with high detection accuracy and short response time. The instrument fully complies with the requirements of the National Pharmacopoeia, and can meet the offline testing requirements of deionized water such as pharmaceutical water (water for injection and purified water) and ultrapure water.

Application

It can be used to detect the concentration of total organic carbon in purified water, water for injection and deionized water in the pharmaceutical industry; it can also be used to detect the TOC of ultrapure water in the semiconductor industry, power plants, scientific research units, laboratories, etc., and is clean in the pharmaceutical and biochemical fields During the verification process, it can be used to verify the cleaning effect.

 

Characteristics

1. No need for nitrogen, oxygen, acid reagents and oxidants, simple operation and low cost of use;

2. It can be used in offline laboratory to meet the test and cleaning verification of organic carbon in water for injection and purified water;

3. Instrument audit trail, simple verification and easy to query;

4. High degree of intelligence, equipped with autosampler, system adaptability verification can be completed automatically;

5. Limit alarm design, when the test sample exceeds the specified limit, the instrument will automatically alarm;

6. It complies with the test plan stipulated in the new edition of the national pharmacopoeia, and can provide IQ/OQ/PQ verification services;

7. Equipped with functions such as electronic signature, audit trail, original data traceability, etc., to meet the verification requirements of GMP, Chinese Pharmacopoeia and 21CFRPART11 computerized system

 

Working principle

 

1. Rotating quartz glass tube coated with titanium dioxide 2. Ultraviolet lamp

3. Conductivity sensor 4. Delay coil 5. Peristaltic pump

 

SPECIFICATIONS

Model

TOC-1000S

Power

100 – 240 VAC, 50HZ, 120W

Measuring range

0.001mg/L~1.5 mg/L (1~1500ppb)

Conductivity detection range

0.055μs/cm~8μs/cm

measurement accuracy

±3%

Resolution

0.001mg/L

RSD

Repeatability≤2%

Sample temperature

0~99°C

noise

<60dB

Detection principle

Using ultraviolet oxidation, direct conductivity principle detection