Thermal conductivity tester (heat flow method), ASTM D5470, MIL-I-49456A

TCM-HF

It is mainly used to test the thermal resistance of thin thermal conductors, solid electrical insulation materials, thermal conductive silicone grease, resin, rubber, beryllium oxide porcelain, alumina porcelain and other materials, as well as the contact thermal resistance at the solid interface and the thermal conductivity of the material. 

More details

Product info

Overview

It is mainly used to test the thermal resistance of thin thermal conductors, solid electrical insulation materials, thermal conductive silicone grease, resin, rubber, beryllium oxide porcelain, alumina porcelain and other materials, as well as the contact thermal resistance at the solid interface and the thermal conductivity of the material. The detection material is in the form of solid sheets, and a frame can be used to detect powdery materials and paste materials.

 

Reference standards

  •      ASTM D5470-2017" Test Standard for Heat Transfer Performance of Thin Thermal Conductive Solid Electrical Insulating Materials", etc.
  •      MIL-I-49456A "Insulating Sheet, Thermal Conductive Resin, Thermal Conductive Glass Fiber Reinforcement";
  •      "Method for Determination of Thermal Conductivity of Beryllium Oxide Porcelain";
  •      "Electrical Insulating Materials" Thermal Conductivity Performance Test Method";

 

Main technical parameters

MODEL

TCM-HF

Sample size

  •       Φ30mm or 20x20mm
  •      Other sizes can be customized

Thermal conductivity test range

  •      0.00150W/m.k
  •      0.001300W/m.k
  •      Range switches by computer software automatically;

Thermal resistance test range

 0.050.000005 M2.K/W;

Sample thickness

  •       0.001~50mm
  •      typical thickness 0.02~20mm;

Thermal pole temperature control range

  •      RT ~ 100°C, RT ~ 299.99°C & RT ~ 500°C is optional
  •      Temperature control accuracy 0.01°C;

Cold pole temperature control range

  •       0~99.00, accuracy 0.01
  •      (equipped with high-precision low-temperature water bath)

Pressure measurement range

  •       0~1000N, accuracy is 0.1N;
  •      Controlled by servo motor, which can accurately set the pressure value of holding pressure

Thickness measurement range

  •       0~50.00mm, accuracy 0.1um
  •      Automatic thickness measurement;

Number of samples

 1 piece (multiple films).

Test accuracy

 better than 3%;

Experimental methods

  •      Thermal resistance test of samples under different pressures,
  •      Material thermal conductivity test
  •      Contact thermal resistance test
  •      Aging reliability test
  •      Material compression performance test
  •      Contact thermal resistance test between samples

Computer & software

Fully automatic computer testing and data printout;

Power supply

 220V/50Hz; 500W

Dimensions (L*W*H)

  •      Host:

550 × 450 × 1555mm (Floor type)

500 × 400 × 785mm (Desktop type)

 

  •      Water Bath:

300 × 350 × 535mm

Net Weight

  •      Host: 100kg (floor type), 60kg (desktop type)
  •      Water bath: 35kg

 

Typical test materials

  •     Metal materials, stainless steel;
  •     Thermal conductive silicone grease;
  •     Thermal conductive silicone pad;
  •     Thermal conductive engineering plastic;
  •     Thermal conductive tape (Sample is very thin and sticky, and it is difficult to make a regular single sample. One side is fixed with transparent plastic and the other side is fixed with paper);
  •     Aluminum substrate, copper clad laminate;
  •     Quartz glass, composite ceramics;
  •     New materials such as copper foam, graphite paper, and graphite sheets.

 

Configuration List

Item name

Qty

Thermal conductivity tester host

1 set

Analysis software (English versions)

1 copy

Precision low-temperature water bath

1 set

Lenovo computer (user can choose);

1 set

Standard samples

2 pcs

Accessories (grease, paste and powder test boxes, Forming stamping mold).

1 set

 