Features:
- Photoelectric detection, digital display, data can be transmitted to the computer to display the melting curve.
The main technical parameters:
- Melting point measurement range: room temperature -300 ° C
- “Starting temperature” setting time: 50°C -300°C no more than 6min
- 300°C-50°C no more than 7min
- Minimum temperature display: 0.1 °C
- Linear heating rate: 0.2 °C / min, 0.5 ° C / min, 1 ° C / min, 1.5 ° C / min, 2 ° C / min,
3°C/min, 4°C/min, 5°C/min
- Linear heating rate error: no more than 10% of the set value
- Indication error: less than 200 °C: ±0.4 °C
- 200 °C -300 °C range: ±0.7 °C
- Repeatability of indication: 0.3 °C at a heating rate of 1.0 °C/min
- Standard capillary size: outer diameter Φ1.4mm inner diameter Φ1.0mm
- Sample filling height: ≥3mm
- Power supply: 220V±22V, 100W, 50Hz
- Instrument size: 398mm x 278mm x 210mm
- Instrument net weight: 12.5kg