PT-D445-005J
Availability date:
Applicable standard
GB/T265 ASTMD445
Performance characteristics
●This instrument has obtained national technology patent, patent number: Z201621447588.6
● Adopt PLC programmable controller system, large-screen LCD touch screen, good interpersonal dialogue screen.
●With rapid temperature rise function, accurate temperature control, fast temperature rise and low thermal inertia.
● Square cylinder, four holes, stainless steel inner tank, front with observation window. With LED lighting.
●The built-in stirring system of the cylinder makes the bath evenly stirred, and the liquid level fluctuation is small; the patented capillary viscometer clamp cover is convenient for loading and unloading, and is stable when used.
● Four capillary viscometers can be measured simultaneously. Touch the test flow time chronograph button to automatically save the chronograph time and automatically calculate the kinematic viscosity value.
● With capillary automatic cleaning system for capillary cleaning.
● If the test temperature is 20 °C, the instrument can be equipped with a refrigeration system, imported compressor, to ensure that 20 °C is not affected by the external temperature to ensure a constant temperature.
Technical indicators
Power supply: AC220V ± 10%
Heating power: 2.0KW
Temperature control range: 20.0 ° C (arbitrary setting)
Temperature control accuracy: ±0.1 °C
Timing: 4 way
Dimensions: Main unit: 740 × 380 × 350 (mm)
Input refrigeration unit: 320 × 3400 × 400 (mm)