Aerosol spray enclosed space tester

CIT10-10S

 CIT10-10S aerosol spray enclosed space test instrument is used to assess flammability of the aerosol in a closed or restricted space. It is in line with the United Nations "on the Recommendations on the Transport of Dangerous Goods • Manual of Tests and Criteria" 31.5 enclosed space ignition test and GB / T 21631-2008 "dangerous sprays enclosed space ignition test methods."

More details

Product info


product description

CIT10-10S aerosol spray enclosed space test instrument is used to assess flammability of the aerosol in a closed or restricted space. It is in line with the United Nations "on the Recommendations on the Transport of Dangerous Goods • Manual of Tests and Criteria" 31.5 enclosed space ignition test and "dangerous sprays enclosed space ignition test methods."

Instrument consists of two parts, the display control section, the detection system composed of a cylindrical container body detection system, built-in light sensors and high-precision thermocouple sensor automatically determine whether knocking occurred. The operator can monitor the entire testing process through the display control section LCD screen.

 

Product Features

  • embedded processor, Windows CE operating system;
  • 8-inch LCD touch screen, display the test status in real time;
  • integration of the "standard model" and "Custom Mode" in two modes;
  • Spray automatic sprays and spray cans clamping device can automatically make the tank vertical and inverted;
  • pressure sensors, nanosecond light sensors double detection module, automatically detects whether a flame is burning or explosion in a closed space;
  • added to the ventilation system, the smoke exhaust from the barrel, to ensure test accuracy and safety of laboratory personnel;
  • testing process fully automated, without human intervention, support parameter setting, data storage, viewing reports, etc;
  • equipped with temperature and humidity module, real-time monitoring the temperature, humidity of test environment .

 

Product parameters

MODEL

CIT10-10S

Working environment

 5 °C ~ 30 °C, <85% RH

Ambient temperature measuring range

 -10 °C ~ + 60 °C

Ambient temp. measurement accuracy

 ± 0.5 °C

Humidity measuring range

 5 ~ 95% RH

Humidity measurement accuracy

 ± 3.0% RH

combustion detection response time

 6ms

Differential pressure measurement range

 -5 ~ 5kPa

Pressure measurement accuracy

 ± 0.1% FS

Timing accuracy

 ± 0.1s

ignition flame

standard candle flame

adjust the pedestal height 20mm / a;

cylindrical vessel size

 Ф600mm × 720mm

aerosol cans Location

 aerosol tank automatically flips

compatible spray cans Dimensions

 H 120 ~ 400mm, D 30 ~ 70mm

Distance from vessel to spray nozzle

 30-35mm adjustable

effusion plate

spray can accumulate discharged residual liquid

It can be withdrawn and processed at any time .

30 other products in the same category: