CIT21-30S
CIT21-30S tester in line with GB / T 21844 -2008, ASTM E681-04 standards, applicable to test the upper and lower limit concentration of combustion measurement of flammable vapor and air mixture formation of chemical substances in testing temperature and constant pressure, it can be used of inert gases as diluents in testing (but the oxidizing of substances should not strong than air)
Availability date:
Product Description
CIT21-30S explosive limit tester in line with ASTM E681-04 standards, applicable to test the upper and lower limit concentration of combustion measurement of flammable vapor and air mixture formation of chemical substances in testing temperature and constant pressure, it can be used of inert gases as diluents in testing (but the oxidizing of substances should not strong than air). During the test, the sample to be tested is first added, heating or cooling the container round to the set temperature, stirring to mix evenly measured gas, high pressure discharge ignition, observe if the test gas explosion combustion through video monitor, if not combustion then adjust the concentration of the measured gas to detect the limit of measured gas.
Product Features
Product Specifications
MODEL | CIT21-30S |
Operating environment | 5 °C ~ 30 °C, <85% RH |
test object | gas or gasified solid, liquid |
Temperature range | -5 °C ~ 180 °C |
Temperature accuracy | ± 1 °C |
Sample temperature measuring range | -50 ~ 300 °C |
The sample temperature measurement resolution | 0.1 °C |
Sample temperature measurement accuracy | ± 1 °C |
Pressure measuring range | 0 ~ 110kPa |
Pressure measurement accuracy | ± 0.1% FS |
The reaction vessel humidity measuring range | 0 ~ 100% RH |
The reaction vessel humidity accuracy | ± 1.0% RH |
ambient temperature measurement range | -10 °C ~ + 60 °C |
Ambient Temperature measurement accuracy | ± 0.5 °C |
Humidity measuring range | 5 ~ 95% RH |
Humidity measurement accuracy | ± 3.0% RH |
ambient atmospheric pressure measurement range | 60 ~ 110kPa |
ambient atmospheric pressure measurement accuracy | ± 0.5kPa |
work to achieve a degree of vacuum | ≤1.33kPa |
Leak degree | ≤0.1kPa / min |
ignition spark gap | 6.4mm |
Ignition Duration | 0.2s ~ 0.4s |
5L glass container dimensions | D222mm, H305mm |
12L glass container dimensions | D295mm, H378mm |
stirrer speed | ≥400rpm |