Impact Sensitivity Tester (Drop Hammer Tester)

CIT18-30S

The test apparatus used to measure energetic materials (solid or liquid explosive, propellants, fireworks and crowded primary explosive, etc.) impact sensitivity, and to determine whether the substance is too dangerous and not under the state testing to transportation.

More details

Product info


Product Description

The test apparatus used to measure energetic materials (solid or liquid explosive, propellants, fireworks and crowded primary explosive, etc.) impact sensitivity, and to determine whether the substance is too dangerous and not under the state testing to transportation.

comply with the United Nations "on the Transport of Dangerous Goods Recommendations on Tests and Criteria Manual" 13.4.2 Test 3 (a) (ii): Federal Materials Inspection Service hammer instrument, "dangerous explosives hit sensitivity Test method "and NY / T1860.6-2010" physical and chemical properties of pesticides test Guideline part 6: explosive ", EN 13631-4: 2002 - explosives for civil uses - potent bomb / drug, part 4: explosives strike the sensitivity of the assay.

 

Product Features 

  • Embedded processor, Windows CE operating system;
  • 8-inch LCD touch screen to ensure display the test status in real time;
  • Use of wireless remote control automatic lifting hammer, hammer positioned to effectively reduce the labor intensity;
  • Wireless remote automatic released hammer, to achieve remote operation;
  • Location automatically engaging hammer designed
  • Unique hammer rebound prevent function, reduce insecurity trials exist;
  • Drop hammer striking head uses special hardened steel materials, sturdy precision, hardness Rockwell 60-63;
  • Unique tooling configuration to ensure accurate positioning of the liquid striking device 2mm;
  • Bottom base fixed with concrete blocks, the whole area of ​​the contact, ensure the stability of the test;
  • Equipped with safety box, effectively protect the safety of personnel;
  • Test selection diversification can configure the test parameters according to the sample characteristics, arbitrary adjustment of the test sequence;
  • Real time display run parameters & state during the test, after the end of the trial automatically analyze the results and gives minimum impact energy of material;
  • Automatic test, support parameter setting, data storage, viewing reports. 

 


Product parameters

Operating environment

 5 °C ~ 30 °C, <85% RH

Impact loading

 0.5kg ± 0.1%, 1kg ± 0.1%, 2kg ± 0.1%, 5kg ± 0.1%, 10kg ± 0.1%

drop height range

 0 ~ 1000mm

Positioning accuracy

 ± 1mm

Rail verticality

 1mm / m

Drop hammer coaxial

 φ 3.0mm

Instrument flatness deviation

 0.20mm

impact energy range

 0.5J ~ 100J

30 other products in the same category: