CIT34-10S
Availability date:
It is suitable for determining the sensitivity of self-reactive substances, organic peroxides and their mixtures to high heat under specified closed conditions.
Comply with the United Nations "Recommendations on the Transport of Dangerous Goods, Manual of Tests and Criteria" 25.4.2 Test E.2 Dutch Pressure Vessel Test and "Safety Code for Inspection of Hazardous Characteristics of Organic Peroxide Dangerous Goods".
Features
1. Using embedded processor, Windows CE operating system
2. 8-inch LCD touch screen, which can display test status in real time
3. Integrate "standard mode", "custom mode" and "calibration mode" to effectively improve test efficiency
4. Remote control ignition, ensure successful ignition through manual observation of the camera, and ensure the safety of personnel operation
5. Configure a flow meter to control the heating rate by adjusting the gas flow.
6. Calibrate the heating rate through a dedicated calibration mode to ensure the accuracy of the test results
7. With self-test and alarm functions, automatic sound and light reminder
8. Configure a video surveillance system to record the test process and observe the final results through playback after the test.
9. The test process is automated without manual intervention and supports parameter setting, data storage, report viewing, etc.
MODEL | CIT34-10S |
working environment | (-5~45)℃,<95%RH |
Heating rate | (3.2~3.8)k/s |
Pressure vessel internal dimensions | Φ50mm×94.5mm |
Pressure vessel material | 316L stainless steel |
Explosion-proof disc material | Aluminum round plate, diameter 38mm |
Explosion-proof disc burst pressure | 620±60kPa @22℃ |
Orifice plate aperture | 1.0mm, 2.0mm, 3.5mm, 6.0mm, 9.0mm, 12.0mm, 16.0mm, 24.0mm |
Orifice plate thickness | 2.0±0.2mm |
Temperature detection resolution | ±0.1℃ |
Temperature sensor response time | 0.5s |