Solid autoignition Point Tester, NF T20-036

CIT22-10S

CIT22-10S solid ignition point tester used to identified spontaneous combustion characteristics through placing the solid material in a steel mesh cube which side length 20mm under heating chamber rising temperature. 

More details

Product info



Product description

CIT22-10S solid ignition point tester used to identified spontaneous combustion characteristics through placing the solid material in a steel mesh cube which side length 20mm under heating chamber rising temperature. It in line with "chemical products for industrial use solids measured relatively spontaneous combustion temperature", EC test A16 and NF T20-036 standards.


Product Features

  • Embedded processor, Windows CE operating system;
  • 8-inch LCD touch screen which can display the test status in real time;
  • Support "standard", "Custom" two modes, taking into account the standards and flexibility tests;
  • Integrated with high temperature insulation technology, temperature control and precise detection technology in one machine.
  • Use electric heating crucible furnace, air circulation natural convection to ensure the sample uniform heating;
  • Furnace chamber materials used vacuum ceramic fiber insulation, good thermal insulation effect;
  • Steel mesh cube hanging at the bottom of the furnace chamber, safe and convenient operation when placing a sample;
  • Automatic heating control, temperature control and combustion testing;
  • Using thermocouple sensor, fast and accurate response to temperature changes;
  • Real time display time temperature graph, raw data is searchable;
  • Testing process fully automated, no need of human intervention, support parameter setting, data storage, viewing reports.

 

Product parameters

MODEL

CIT22-10S

Operating environment

5 °C ~ 30 °C, <85% RH

Test object

Solid

Temperature range

room temperature ~ 650 °C

Temperature accuracy

Below 350 °C of ± 1 °C

Above 350 °C of ± 2 °C

Temperature measuring range

0 ~ 800 °C

Temperature measurement resolution

0.1 °C

Temperature measurement accuracy

Below 300 °C of ± 2.5 °C

Above 300 °C of ± (0.75% of reading)

Heating Rate

0.5 °C / min

the heating chamber capacity

2L

Timing range

0-24h

Timing accuracy

± 1% Timing range

Cage size

20 mm side length, aperture 0.045㎜