Full Automatic closed flash point tester

PT-D93-002ZII

Product info

Applicable standard
GB/T 261-2008, ASTM D93 (A and B methods)

Performance characteristics

  1. PLC automatic detection, control, temperature programming, automatic ignition and judgment of flash point, test results can be stored, printed, downloaded, the entire test process without the need for personnel participation.
  2. 7-inch color LCD touch screen, full Chinese operation interface.
  3. With automatic lifting lid function, easy to operate.
  4. The stirring device is designed by the national patent. The stepping motor is used for sample stirring, and the rotation speed is stable and adjustable. It can be tested by A and B methods.
  5. Built-in atmospheric pressure sensor can automatically correct the measurement result according to the measured atmospheric pressure.
  6. with automatic cooling to improve cooling efficiency <. Depending on the user's needs, the software and interface to connect to the LIMS system can be configured.

Technical indicators
1 power supply: AC220V±10% 50Hz
2 Measurement range: 40 ~ 300 ° C
3 sensor accuracy: 2‰
4 flash point detection: miniature thermocouple
5 heating rate:
Method A: 5 to 6 ° C / min
Method B: 1 to 1.5 ° C / min
6 heater power: 600W
7 stirring speed: Method A: 90 ~ 120r / min, Method B: 240 ~ 260r / min
8 gas source: liquefied petroleum gas, propane

  • Dimensions: 240 × 410 × 39 (mm)