GC2002
Availability date:
Shanghai Science and Technology "15" scientific and technological projects, the country's first color touch screen operation
Instrument Features:
1, The touch screen with a durable, fast response, space-saving, easy to communicate, long life; there are vivid, lifelike and rich images and interact with us.
2, Whole computer control systems, computer recriminations, a computer can control the N (N ≤ 253) Chromatography with a six-way temperature control system; four way time program system.
3,The use of advanced technology Chromatography 10 / 100M Ethernet communication interface enables the instrument remote control and remote data transfer process and supervision. It may be connected to the unit supervisor and up leader, to facilitate the implementation of competent supervision.
4, You can select the analog signal output, also choose the digital signal output (built-in Chromatography workstation), signal cable direct output.
5, You may choose two or three kinds of commonly used detector at the same time.
6, Large capacity oven with automatic rear door, can be 8-level temperature programming; near room temperature control function (above room temperature 8 ℃).
7, Can be configured packed column injector, a capillary column injector, a gas injector, conversion path thermal desorption apparatus, headspace, thermal cracker, auto sampler and the like.
8, With fault self-diagnosis function, display fault location and nature at any time; over-temperature protection, temperature exceeds the set temperature in any way, it will automatically stop heating.
Technical Parameters:
Thermal Conductivity Detector | Flame detector (FID) | Electron capture detector (ECD) | Flame photometric detector (FPD) | |||
TCD-1 | TCD-1H | MTCD | FID |
|
| |
Sensitivity (detection limit) | ≥3000-10000mV.ml/mg(benzene) | ≤5.0×10-12g/s(N-hexadecane) | 1×10-13g/ml(γ-666) | 5×10-12g/s(P) | ||
Baseline noise | ≤0.01mv | ≤1×10-13A | ≤40 uV | ≤40 uV | ||
Baseline drift | ≤0.1mv/30min | ≤ 5×10-13A/30min | ≤100uv/30min | ≤100uv/30min | ||
Linear range | 105 | 107 | 104 | 104 | ||
Settling Time | ≤1.5h | ≤1h | ≤2h | ≤1h |
2, Temperature control
Temperature range: Oven volume: 280 × 280 × 180mm
Temperature range: room temperature above 5 ℃ -400 ℃
Temperature Accuracy: ± 0.5%
Display accuracy: 0.1 ℃
Program level: 8 level
Cheng liter rate: 0-39 ℃ / min (adjusted increments 0.1 ℃ / min)
Programmed temperature Repeatability: ≤1%
3, Host size: 650 × 480 × 500mm
4, Instrument Weight: 53kg
5, Power: 2000W