GC9800
Availability date:
Won domestic good instrument honors
New GC9800 Network Gas Chromatography (hereinafter referred to as GC9800 (N)) is our own research and development of a new generation of high-performance intelligent network Gas Chromatography, compared to older generation products, with rapid oven cooling, high sensitivity, LCD display, built-in Chromatography workstation, reversible control operations and network remote supervision six advantages. It is widely used in areas such as in oil and gas, petrochemicals, fine chemicals, coal chemicals, gas, food, medicine, agriculture, environmental protection, scientific research to do the mixture constant component and a trace component analysis.
Chromatographyic detection signal of GC9800 (N) gas Chromatography has analog and digital outputs two kinds, the user can choose one as needed. Analog signal output can be connected with any Chromatography processor, and any Chromatography workstation; digital signal output is connected through a network interface to a PC with a built-in native Chromatography workstation for Chromatography data acquisition and analysis of data in real-time control and anti-Chromatography run. A computer can be connected with two hundred and fifty Chromatographys, operate and control of Chromatographys and obtain Chromatographyic data.
GC9800 (N) gas Chromatography is an intelligent network Chromatography instrument has six-way thermostat and three-way time process control, instrumentation heating fault automatic alarm and alerts. After mounting HT300A auto sampler, up to 110 liquid samples in the case of unattended, for automatic sample analysis. Operational status of the instrument networking capabilities of the instrument can result samples analyzed by real-time transmission network to the enterprise and industry department in charge of leading sectors. Business managers at all levels to facilitate the timely analysis of data, data processing and analysis, three-level network monitoring.
Main Feature:
1, Whole computer control systems, computer recriminations, a computer can control the N (N ≤ 253) Chromatography with a six-way temperature control system; four way time program system.
2,The use of advanced technology Chromatography 10 / 100M Ethernet communication interface enables the instrument remote control and remote data transfer process and supervision. May be connected to the unit supervisor and up leader, to facilitate the implementation of competent supervision.
3, You can select the analog signal output, also choose the digital signal output (built-in Chromatography workstation), signal cable direct output.
4, You may choose two kinds of commonly used detector at the same time. (FID, TCD, ECD, FPD 1-2 selected species)
5, Large capacity oven with automatic rear door, can be 8-level temperature programming; near room temperature control function (above room temperature 8 ℃).
6, Can be configured packed column injector, a capillary column injector, a gas injector, conversion path thermal desorption apparatus, headspace, thermal cracker, auto sampler and the like.
7, With fault self-diagnosis function, display fault location and nature at any time; over-temperature protection, temperature exceeds the set temperature in any way, it will automatically stop heating.
Technical Parameters:
1.Detector indicator
| Thermal conductivity detector (TCD) | Flame detector (FID) | Electron capture detector (ECD) | Flame photometric detector (FPD) |
Sensitivity (detection limit) | ≥5000mv.mL / mg (benzene) | ≤5.0×10-12g/s | 1×10-13g/ml(γ-666) | ≤5×10-12g/s(P) |
Baseline noise | ≤10uv | ≤1×10-13A | ≤40 uV | ≤40 uV |
Baseline drift | ≤60uv/30min | ≤ 5×10-13A/30min | ≤100uv/30min | ≤100uv/30min |
Linear range | 105 | 107 | 104 | 104 |
Settling Time | ≤2h | ≤1h | ≤2h | ≤1h |
2, Temperature control
Temperature range: Oven volume: 250 × 250 × 180mm
Temperature range: room temperature above 8 ℃ -400 ℃
Temperature Accuracy: ± 0.5%
Display accuracy: 0.1 ℃
Program level: 8 level
Cheng liter rate: 0-39 ℃ / min (adjusted increments 0.1 ℃ / min)
Programmed temperature Repeatability: ≤1%
3, Instrument under normal operating conditions
Ambient temperature: 5-35 ℃
Relative humidity: ≤85%
Power supply voltage: 220V ± 10%, 50Hz ± 0.5Hz;
Indoor non-corrosive gases, the table may not have a strong vibration, strong magnetic field around does not exist.
4, Host size: 500 × 540 × 480mm
5, Power: 1500W