As we all know, nano-materials science and engineering has become a worldwide hot topic in the study of surface modification of nano-materials, often involves wetting contact angle of this concept. The so-called contact angle refers to the drop on a solid horizontal plane a droplet on a solid surface solid - liquid - gas-phase junction point, the gas - liquid interface and solid - liquid interface of the two tangents when the folder in which the liquid phase the angle formed. SI-CAM2000Y series is to measure the uneven surface contact angle measuring instrument, mainly for hard glasses, contact lenses industry, glass and other circular pipe industry. Japan HOYA optical procurement is this series of instruments.

Features and technical parameters 
1. Instrument composition
basic configuration:
Peristaltic sample pump, video capture card, manual tilt platform CCD, CCD camera, continuously variable power optical system (domestic), fixed sample platform, uneven special image analysis measurement system
Optional:
Manual ± 15 ° swing arc sample platform, manually rotating the sample platform 360 ° vertical, 360 ° horizontal rotation manually sample platform, automatic electric temperature control system (Omron temperature control), continuously variable power optical system (Nissan), automatic precision approach Thus, automatic image analysis system, all-electric three-dimensional platforms, customized special convex surface of the base

2. Main function:
Measuring the contact angle (measured uneven surface dedicated software)
Measurement of the surface tension (hanging drop method)
Observed and measured droplet shape and size (hanging drop method)
Calculate the surface free energy (Owens two-calculation method)
3. Measurement method:
      Flash frozen screenshot
      Continuous automatic shooting pictures (1-3600 seconds to shoot a picture)
      Automatic high-speed photography (12 frames per second)
4. Contact angle measurement range: 0-180 °
5. Image magnification: 55pixel / mm ~ 315pixel / mm
Imports continuously variable power optical system: (100pixel / mm ~ 1400pixel / mm) SI-CAM2000Y contact angle meter
6.Accuracy: 0.1 ° or 0.5 ° (convex surface special type)

Optional
1, a semiconductor temperature control platform (using Omron temperature controller, controllable heating, room temperature ~ 98 ℃)
2, advanced zoom optical system and CCD camera (imported from Japan)
3, automatic precise loading sample
4, customized special convex surface of the base
5, manual vertical arc put the sample platform (plus or minus 15 degrees)
6, manual sample platform 360-degree vertical rotation
7, 360-degree horizontal rotation Manual sample platform
8, all-electric three-dimensional platform (sample platform, focusing)
9, 10ul / 5ul micro-injector, disposable micro-injector and fixture
Note: 1,5,6,7 four platforms can simultaneously buy at option, but can’t use together.