PRE-DG Series
Availability date:
Application
Liquid chromatography coupled, which are widely used in biological pharmaceutical, fine chemicals and polymer drug analysis, environment analysis, analysis, proteomics, forensic toxicological analysis, food testing and chemical testing, etc.
Introduction
In high performance liquid chromatography (HPLC) analysis system, if the mobile phase of the carried sample containing bubbles, abnormal peak will appear .Seriously affect the detection precision of the detector.Although degassing of the mobile phase during pretreatment, but due to the need of mobile phase chemical properties and the analysis itself, when two mixed solution will produce bubbles.Online degasser can continuously remove gases from liquid phase , eliminate the unstable factors of the mobile phase and lower the baseline drift and noise, thus eliminated interference to detect oxygen by the electric chemical , fluorescence and uv detection .
HDG series online degassing instrument adopts the advanced gas/liquid membrane separation technology, chooses import original sensor, can automatically start the vacuum pump in order to guarantee the degassing effect.Standard joint and make it a pipe connected to various types of HPLC system.
Working Principle
HDG series degassing instrument is using the principle of vacuum degassing.When the mobile phase in liquid pump driven by placed in a vacuum chamber to take off the trachea, tube inside and outside the general pressure makes the original mobile phase in the bubble molecules from continuous precipitation separation membrane.Length in the vacuum chamber to take off the trachea design can satisfy the requirement of the conventional traffic degassing.
Main Features
1. The advanced gas/liquid membrane separation technology, the degassing efficiency is high, can meet the strict requirements of liquid chromatographic analysis system.
2. The degassing system all closed, not change the chemical composition of the mobile phase.For flammable, volatile liquid degassing, can guarantee it will not overflow, so as to achieve the security of the degassing process.
3. The installation is simple, easy to operate.The device of automatic startup system makes operator on line at a time, can be continuous use.
4. Due to the international standard connectors and pipe, make it apply to all kinds and any manufacturer of high performance liquid system.
5. Complete specifications, produced a series HDG degassing instrument , 1-4 channels, high cost performance.We also provide different requirements of customers import original qi activity.
Item No, | HDG-02 | HDG-03 | HDG-04 |
Channel | 2 channel | 3 channel | 4 channel |
The degassing efficiency | Oxygen 8 PPM solution at 25 ℃ import, export solution oxygen: 1.8 PPM when the flow rate of 1 ml/min, flow rate of 5 ml/min 4 PPM. | ||
Vacuum pressure Mpa | 0.085 | ||
Maximum flow ml/min | 10 | ||
Amount of solvent (each way) | 12ml | ||
Overall dimensions mm | 150×340×180 |