-10°C~200°C Industry Hermetic Refrigerating & Heating Circulator (Dynamic Temperature Control systems)

HRHC Series

Application:
Reactor Systems Autoclaves, Pilot Systems, Mini Plant Systems, Scale up for Operational Development, Double wall reactors, Advanced-Flow Reactors, Distillation Systems, Material Testing, Combinational Chemistry, Semi conductor Industry, Vacuum Chambers.

More details

Product info

Product Video:https://youtu.be/Tfn6DHmV9yY

Advantages & Functions

  • Hermetic circulation system make sure high accuracy of temperature and extend life time of circulating liquid.
  • Widely working temperatures range from -100°C to +350°Cwithout cooling fluid change
  • Intelligent PID temperature control
  • High stability and reproducibility of process
  • Adopted plate heat exchanger, heating pipeline to improve rate of heating & cooling
  • Widely cooling power range from 0.5 to 500 kW
  • Brilliant 4 inch, 7 inch, 10 inch TFT touch screen with graphic display can be choose.
  • Comprehensive warning and safety functions
  • Rapid cooling down from high temperature (from 300°C).
  • Control software is available which can communication in 200 meter distance. (Optional)
  • Incredibly compact

Structure design

A fully closed piping design, high efficiency plate heat exchanger used to reduce the demand for the thermal fluid utilization while providing the system to achieve the rapid ups and downs in temperature The whole system is a closed liquid circulation system with a expansion of container, expansion of the container and the liquid circulation is adiabatic, it will not in liquid circulation, only mechanical connection, regardless of the temperature is a high temperature or low temperature, the tank temperature is limited to 60°C. The advantages by this design:

  • Only the heat conduction medium in the expansion container touches the oxygen in the air. (The container temperature is limited to 60°C), therefore, during the process the risk of imbibing and volatilization of the heat conducting medium is reduced.
  • No volatilization of the heat conducting medium at high temperature.
  • No need to change the heat conducting medium. Continuous control the temperature range to -80°C~190°C, -70°C~220°C, -88°C~170°C, -55°C~250°C, -30°C~300°C without rising the pressure.

 

Display function

  • All the kind of process control temperatures.
  • Liquid level of the heat conducting medium in the expansion vessel.
  • Indication working status of refrigeration system.
  • Indication working status of heater.
  • Indication working status of circulation pump.
  • Temperature control mode (materials and heat transfer medium).
  • Temperature upper limit and lower limit can be set.
  • Temperature difference between jacket & reactor material can be set.
  • Alarm when liquid is empty.
  • Refrigeration compressor can be set to operate manually or automatically.
  • Clear temperature curve displayed, it can be zoom and exported to computer.
  • Recipe procedure set display
  • Alarm note display

Principle of process control
Change the control settings of the method, the response as soon as possible in the process of the system lags behind, be the smallest system overshoot. Controlled by two PID (PID is a variable in each group) control loop structure, known as the two sets of control loops: the main circuit and from the circuit, the main control loop from the loop output as the settings. System with feed forward with PV, the master PID loop operation results of the PV output and feed forward signals  as the composite of the settings from the control loop, through such control of the temperature gradient to ensure the accuracy of temperature control system.

  

Circulation Pump
High temperature resistant low tempera  ture magnetic drive pump or a leak-free high and low temperature shield-driven pump, reduce the risk of system leakage. At the same time the system has the advantages of low noise and big traffic flow.

Electronic expansion valve
Wide temperature range of dynamic temperature control system with electronic expansion valve produced by Emerson company, It can monitor and adjust the number of steps (0~750 Steps / minute) into the motor, precise control the liquid volume from the refrigerant , Improve control accuracy.

Configuration software (Optional)

All dynamic temperature control system HRHC-Series can be connected with the computer.
You can install and record the picture of display on the computer via the configuration software.

  • Communication distance less than 200m.
  • Easily temperature setting
  • Real time control screen
  • Curve Records
  • Program selection
  • Alarm screen recording

 

Safety protection
Many additional features to ensure the safe operation of the system. For the operation to run not normally appear in the alarm in time and displayed on the touch screen. Phase sequence, leakage protection, refrigeration system high and low voltage protection, overload the pump, compressor overload, high and low level, two independent over-temperature protection, water protection, protection of the circulation line shut.

Connecting Pipe


Items

Temp. range

Interface port

Fluorine rubber hose

-30~200°C

Φ12*16  Φ16*22  Φ20* 26

Metal insulation pipe

-60~250°C

DN15  DN20  DN25  M24*1.5  M30*1.5  M38*1.5

 

Date interface & software
Standard configuration:

  • Inputs of PT100 temperature sensor
  • USB data export interfaces
  • RS485 interface (MODBUS protocol)
  • Alarm contact output

Optional:

  • Inputs of 4 ~ 20mA temperature measurement
  • Inputs of 4 ~ 20mA temperature setting
  • Ethernet interface
  • RS232 interface (MODBUS protocol)
  • Computer operating software (Must be fitted with Ethernet interface)
  • External control box with touch screen display

Common Specifications:

  • Control mode: Feed forward PID, PLC control
  • Temperature control: Temperature of material and circulation liquid can be controlled and set
  • Program:20 group programs, 45 segments for each group
  • Protocol: MODBUS RTU / RS 485 Interface
  • Temp. sensor: Circulation medium PT100, raw material PT100 (3 measuring points) 4~20mA (Optional)
  • Control panel: 7 inches color touch screen, temperature curve display & record
  • Safety: With self-diagnosis function; freezer overload protection; high-pressure switch, a variety of security features overload relays, thermal protection devices.
  • Closed loop system: Fully closed system, no oil mist at high temperature, low temperature do not absorb moisture in the air, the system does not increase the pressure because the high temperature, low temperature heat transfer medium in automatic replenishment operation.

-10°C~200°C Industry


Model

HRHC-1A25W

HRHC-1A38W

HRHC-1A60W

HRHC-1A95W

HRHC-1A130W

Temperature range

-10°C ~ 200°C

Material temp. accuracy

±1°C

±1°C

±1°C

±1°C

±1°C

Cooling power

200°C

25kW

38kW

60kW

95kW

130kW

0°C

25kW

38kW

60kW

95kW

130kW

-10°C

15kW

22.8kW

36 kW

57kW

78kW

Circulation flow rate

150L/min
2.5 bar

250L/min
2.5 bar

250L/min
2.5 bar

250L/min
2.5 bar

400L/min
2.5BAR

Compressor

Copeland

Copeland

Copeland

Copeland

Copeland

Expansion valve

Emerson

Emerson

Emerson

Emerson

Emerson

Evaporator

Shell-tube evaporator

Refrigerant

R-404A

Port size

DN-25PN10

DN32PN10

DN40PN10

DN40PN10

DN40PN10

Rate & pressure (water-cooled)

8m3/h
1.5bar~4bar.

10m3/h
1.5bar~4bar

15m3/h
1.5bar~4bar

20m3/h
1.5bar~4bar

28m3/h
1.5bar~4bar

Dimensions

1250×700×1500

1500×800×1600

1800×800×1650

2000×1000×1650

2300×1250×1750

Weight

640 kg

900 kg

1150 kg

1400 kg

700 kg

Power

AC380V 50HZ
32kW (max)

AC380V 50HZ
48kW (max)

AC380V 50HZ
75kW (max)

AC380V 50HZ
125kW (max)

AC380V 50HZ
155kW (max)

External material

SUS 304

SUS 304

SUS 304

SUS 304

SUS 304

Power

440V~480V 60Hz(Single or three phase) is optional

26 other products in the same category: