CB-H series
Availability date:
Application Range:
The clean bench is applicable in medicine, pharmaceutical, chemical experiments, electronics, defense, precision instruments, meters and other industries.
Structural Characteristics:
Vertical Lamina Flow Specifications:
Model Item | CB-H1F | CB-H2F | CB-V1300FD | |
Clean Level | Class 100@≥0.5um(U.S.Federal209E) | |||
Average Air Velocity | 0.32~0.66m/s(Recommendedspeed0.3m/s) | |||
Colonies | ≤0.5 units/dish.hour(φ90mmPetri dish) | |||
Noise | ≤62dB(A) | |||
Vibration Half Peak | ≤5um | ≤5um | ≤5um | |
Luminance | ≥300Lx | |||
Electric Power | AC Single-phase 220V/50Hz | |||
Maximum Power | 0.4Kw | 0.8Kw | 0.8Kw | |
N.W. / G. W.(Kg) | 80/130 | 130/210 | 130/210 | |
Size | W1: Workspace W | 87 | 136 | 112 |
D1: Workspace D | 70 | 70 | 65 | |
H1: Workspace H | 52 | 52 | 72 | |
W: Outer W | 99 | 148 | 136 | |
D: Outer D | 72 | 72 | 80 | |
H: Outer H | 166 | 166 | 180 | |
Package Size(cm) | 114×86×186 | 164×86×186 | 144×94×191 | |
HEPA Filter Dimensions & Qty. | 820×600×69×(1) | 1310×600×50×(1) | 610×610×69×(2) | |
Number of Operators, Work Area | 1F: one person two sides | 2F: two person two sides | two person one side | |
Notes | U means UV Light |