BCA-4060
Availability date:
Overall Performance
Equipment type Fully automatic discrete, STAT priority
Analysis rate Constant speed 460T/H (single/Double reagents, without ISE)
Test principle Colorimetry, turbidimetry
Analysis method End-point, 2 kinetics, fixed-time, etc. support single/double wavelength, linear and non-linear calibration
Sample System
Sample Tray 90 sample positions, random access of all samples
Sample Volume 2-50ul, 0.1ul increment
Sample probe Liquid level detection, the remaining detection, collision protection etc.
Sample Containers Sample cups, Micro tube, Original blood-collection tube, etc.
Reagent System
Reagent Tray 90reagent positions, two trays
Reagent Volume R1:25-300ul, R2:10-150ul, 0.1ul increment
Reagent Probe Liquid level detection, the remaining detection, collision protection etc.
Reagent Refrigeration 4-15℃, independent power supply, 24-hour non-stop refrigeration
Reaction System
Reaction Tray Rotating tray, 120 cuvettes
Cuvette Reusable high quality cuvette, Optical diameter 6mm
Reaction Volume 180-500ul
Incubation system Bi-directional liquid-air device
Temperature 37℃±0.1℃ with real-time display
Optical System
Light source 12V35W halogen lamps, over 2000 hours
Photoelectron road after spectrophotometry
Wavelength 340,405,450,510,546,578,630,700nm
Absorbance range 0-3.500Abs
Resolution 0.0001Abs
Operating System
PC operating system Windows XP or above
Analysis control software English version graphical operating software
Report printing Report formats support the user-defined mode, QC and state information etc.
System connection RS-232
Working Conditions
Power supply AC 220V±22V, 50Hz±1Hz, power 1.5KVA
Ambient Temperature 15-30 ℃
Humidity 35-80%
Water consumption 8L/Hour
Dimension 950mm*800mm*1100mm
Weight 140KG