PKT-N530
PKT-N530 Air Permeability Analyzer can tests the O2, CO2, N2 etc. gas transmission rate of plastic film, composite film, aluminum foil, thin sheets, rubber, porcelain, solar panel, medical patch and so on.
Availability date:
PKT-N530 Air Permeability Analyzer can tests the O2, CO2, N2 etc. gas transmission rate of plastic film, composite film, aluminum foil, thin sheets, rubber, porcelain, solar panel, medical patch and so on.
Working Principle
Differential-pressure method. Test sample separates test chamber into upper room and lower room, between which keeps a constant pressure difference (100KPa). Gas molecules penetrate through sample from higher pressure room into lower pressure room; measure the pressure change in lower pressure room and calculate gas transmission rate.
Standards
GB/T 1038, ISO 2556, ISO 15105-1, ASTM D1434
Features
Software Interface
Specifications
Test range | 0.02~50000 cm3/m2·24h·0.1MPa |
Test precision | 0.01 cm3/m2·24h·0.1MPa |
Temperature range | 15~35℃ (5~50℃ optional) |
Temperature precision | ±0.01℃ |
Vacuum degree | <10Pa |
Vacuum resolution | 0.1Pa |
Test pressure | 0~0.2MPa |
Gas source | test gas, purity 99.999%, 0.2~0.8MPa |
Gas port | Φ6mm gas pipe |
Test gas | O2, CO2, N2 etc. |
Test area | 50.24 cm2 |
Sample size | Φ110 mm |
Sample thickness | ≤2mm |
Number of samples | 1~3pcs |
Instrument size | 80×56×40cm |
Configuration
Air Permeability Analyzer, software, communication cable, power cable, sample cutter, gas pipes 2pcs, pipe connectors 2pcs, sealing grease, spanner, filter paper 9pcs, Standard Films 1 set
Optional: vacuum pump, pressure releasing valve, computer
User provide: gas source
sample cutter vacuum pump pressure releasing valve