Basic functions:
1. Touch screen control, automatic self-checking on starting-up;
2. Protection alarm of water shortage and power failure;
3. Built-in RO membrane fouling prevention program, RO membrane-wash keys;
4. IC Microcomputer intelligent control;
5. Built-in automatic timing cycle process of ultrapure water, which maintains high water quality;
6. Alarm of consumable goods expiration;
7. On-line monitoring of water quality resistance;
8. 128×64 large color LCD screen;
9. Display of running condition and running parameters;
10. Function of water quality and quantity upgrading;
Technical parameters:
Application scope |
DNA / RNA preparation; DNA sequencing; HPLC (amino acids, penicillin analysis; organic component testing); general immunology; microbial applications; electrophoresis; PCR; immune diagnosis; protein purification, protein analysis; toxicity; TOC detection; |
||
Mode |
S |
D |
V |
Description |
Single RO |
Double RO |
RO water or distilled water as water source |
(Production rate) |
Molgene 1805a 5L/H |
Molgene 1805d 5L/H |
Molgene 1805V |
Water source |
City tap water, water pressure1.0-5.0kg/cm2 |
Continuous water flow |
1.5L/min; |
Outlet 1 |
Resistance of III-class pure water is 0-10us/cm, conforming to GB6682-92 of III-class water standards and ASTM CAP NCCLS standards; |
Outlet 2 |
Resistance of I-class ultra-pure water is up to 18.25 MΩ.cm 25 °C, conforming to GB6682-92 standards of I-class water and ASTM CAP NCCLS standards; |
Retention rate |
Si> 99.9%; |
Absorbance (254nm, 1cm light path) |
≤ 0.001; |
Soluble silicon (SiO2) |
<0.01ppm; |
Particles content (> 0.22μm) |
<1/ml; |
Heavy metals (ppm.max) |
<0.001; |
TOC |
< 1~3 ppb |
Terminal microfiltration |
UF 6000 dalton |
Microorganism |
< 1 cfu/ml |
UV |
254 nm & 185 nm |
Resistance |
18.25MΩ.cm |
heat source |
< 0.001 EU/ml |
Power Supply |
220V/50Hz, voltage fluctuation rate of ± 10%; |
Working conditions |
temperature 5-40 °C, anti-freezing rooms are needed in cold areas; |
Dimensions |
460 × 480 × 500 (mm). |